UCS20 USB Computer Spectrometer
Click on photo to enlarge
UCS20 Universal Computer
Spectrometer
Note: The UCS20 has been replaced by our NEW UCS30.
Click here to go to
the NEW UCS30 page.
Featuring a maximum of 2048 channels, this fully integrated Plug and
Play Multichannel Analyzer connects to the USB port on a PC or Macintosh
eliminating the need to install a card inside the computer. The
instrument includes an internal preamplifier, computer controlled
amplifier, 0-2500V HV supply, LL and UL discriminators, PHA, MCS and
Mossbauer modes and full operating software.
Designed for flexibility to cover a wide range of applications,
the UCS20 will support NaI(Tl) detectors, CdZnTe and proportional
counters for gamma and X-ray spectroscopy. Preamplifier power output
(±12V) for devices requiring
external preamplifiers and direct inputs to the internal amplifier and
ADC are included with external gating capability for coincidence
measurements.
In addition to the Pulse
Height Analysis mode, the UCS20 also includes Multichannel Scaling for
time related studies. Dwell times as low as 100msec. combined with a MSB output and
multi-pass capability provide interfacing to many Mossbauer
systems.
The software supplied is
compatible with Window 98, ME, 2000 and XP and Macintosh System 8.6
through 9 supporting USB. All setup controls including amplifier gain,
HV, LLD, ULD, Conversion Gain, Preset Time, Preset Count and Mode are
accessible from the software. Data is stored in binary format or as TSV,
CSV format for direct transfer to spreadsheets. A data buffer is included
for storing a background spectrum that may be fractionally subtracted
from the acquired data if required.
To assist in initial setup,
an Auto Calibrate feature is included to provide a linear calibration
using Cs-137 of 1 keV/channel at 1024 channel conversion gain. Alternate
setups for specific measurements may be save as Setup files for quick
reconfiguration of the system. All current operating parameters are saved
so the application program may be exited and reentered without disturbing
acquisition or loss of data. Once operational, the UCS-20 will run
autonomous of the computer as all data and control parameters are stored
in the instrument providing power is maintained.
To aid in the identification of nuclides, the UCS-20 contains a unique
peak labeling feature named ISOMATCH. Providing the unit is accurately
energy calibrated, the user may select a nuclide from the library and the
corresponding characteristic emission lines are superimposed on the
spectrum along with isotope and energy information. Using this feature,
users may quickly check a spectrum and visually identify the emission
components for each nuclide present. The ISOMATCH libraries may be
created or customized using a text editor.
Rear Panel
Specifications.
Hardware:
- Physical: Fully shielded bench top metal enclosure, 9x13x2.5
in.
- Preamp In: BNC connector for direct connection to PMT anode
signal. Utilizes on-board amplifier/preamplifier.
- Amp In: BNC connector bypasses internal preamplifier for use
with external preamplifier. Accepts positive polarity signals.
- Direct In: BNC connector bypasses internal amplifier and
connects directly to ADC input. Accepts unipolar or bipolar signals
0-8volt range.
- Gate In: BNC connector. Normally open gate. TTL low closes
gate.
- Ext MCS In: BNC connector bypasses internal amplifier.
Accepts positive TTL signals >150nsec., 5MHz maximum rate.
- HV Output: MHV (or optional SHV) connector supplies positive
0-2500volt @ 1mA maximum current to power scintillation detector High
voltage is fully regulated and computer controlled in 5volt
increments.
- Preamp Power: DB-9 connector, ±12V.
- MSB Out: TTL output of dwell MSB for synchronizing Mossbauer
drive.
- ROI Out: TTL pulse each time event occurs in set region of
interest.
- USB: Standard USB 2.0, type “A” connector.
- Amplifier: On-board combination preamp./amplifier for use
with scintillation detectors and PMT's. Computer controlled coarse and
fine gain from x2 to x1000.
- ADC: Wilkinson type with 80MHz clock and computer selected
conversion gain of 2048, 1024, 512, or 256 channels. Direct input
accepts pulse peaking times of 1sec. to 10sec. Includes deadtime
correction when used in Live-Time mode.
- LLD & ULD: Independently computer controlled in 4-channel
increments over entire input range. Operates prior to ADC for reduced
system dead time.
- Modes: MCA for pulse height analysis, or MCS for half-life
decay or other time related studies.
- Timers: Real-Time or Live-Time operation selectable from
1sec. increments for PHA and 100msec. to 60sec. per channel in MCS
mode.
- Data Memory: On-board dynamic RAM, 1 x 2048 channel x 3 Bytes
for data, plus region-of-interest flag.
- Deadtime: System dead-time is computed and displayed on
screen during acquisition.
- Power: AC line, 100-250VAC, 50/60Hz, auto-sensing
internal power supply with Universal Power Cord input.
- Status Indicators: High voltage, Acquire, Activity.
Software.
- System: Operates under Window 98 and higher or Macintosh
System 8.6 and higher.
- Display: Color or monochrome monitor, Vertical scale adjust
from 32 to 16M and LOG display. Horizontal 2048 channels with
expansion down to 128 channels.
- Time Mode: Preset live-time or preset real-time selection.
Both times are recorded and displayed.
- Count Mode: Preset count mode integral of
region-of-Interest.
- ROIs: Multiple Regions-of Interest using color coding.
- Integral: When cursor is in ROI, computes gross area, net
area with end point averaging, centroid and FWHM.
- Energy Cal: 2-point linear or 3-point quadratic converts
cursor position reading directly to energy units. (Time units in MCS
mode.)
- Autocalibrate: Use with Cs-137 source to automatically
calibrate system from 0-1023keV.
- Bkgd Buffer: Used to store background spectrum for time
normalized background subtraction.
- Subtract: Subtracts channel-by-channel time normalized data
stored in Background Buffer from display data.
- Smooth: 3-point smoothing of displayed data. Does not modify
data in acquisition memory.
- Control: Software control of High Voltage, Amplifier Gain,
Lower and Upper level discriminators, and ADC Conversion Gain.
- File: Save or load data file and header information in binary
format or as CSV/TSV files. Save or load setup files for specific
configurations.
- Clipboard: Clips data and header information, graphic
screens, ROI data containing, centroid, FWHM, gross and net integrals,
and header information in spreadsheet compatible format for
transferring into other applications.
- ISOMATCH: Isotope library text file with peak markers and
labeling for overlaying on spectrum. Provides quick isotope
identification. Libraries may be edited and extended using word
processor or text editor programs.
Ordering Information.
- UCS 20 Universal Computer Spectrometer for PC
and Macintosh.
Options
- LS325 Environmental lead shield for 2" or 3"
detectors.
- SD51 2x2 in. NaI(Tl) with integral photomultiplier.
- TS51-2 Tube base stand, 14-pin for 2" or 3" detectors,
complete with cables.
- ST100 External preamplifier for proportional
counters.
- SP78W-2 Well detector complete with stand, lead shield and
cables.
- SPA 38 Complete detector assembly with 38x38mm detector,
stand, shield and cables.
|